Auto Hematology Analyzer BK-5000
Model
|
BK-5000
|
Automatic
|
Fully automatic
|
Throughput
|
60 tests/hr
|
AssayItems
|
22 parameters, 3 parts differential of WBC
|
MeasuringPrinciple
|
Electro-impedance, photometric analysis
|
Testing Items
|
WBC LY LY% MO MO% GR GR%
RBC HGB HCT MCV RDW-SD RDW-CV MCH MCHC PLT MPV PCT PDW WBC histograms RBC histograms PLT histograms |
LinearityRange
|
WBC: 0.0-99.9(×10^9/L) RBC:0.0-99.9(×10^12/L)
HGB: 0.0-300(g/L) PLT: 0.0-99.9(×10^9/L) |
Accuracy (CV%)
|
WBC≤2.0% RBC≤1.5% PCT≤3.0%
PLT≤5.0% HGB≤1.5% MPV≤3.0% HCT≤2.0% MCV≤1.0% PDW≤3.0% MCH≤2.0% MCHC≤2.0% |
Display
|
LCD
|
Storage
|
200,000 results including histograms
|
Output
|
One RS232 ,one PS2 and two USB. Available for LIS
|
Printer
|
Built-in thermal printer, USB to external printer(Optional)
|
Data input
|
Keypad ( mouse and keypad available)
|
QC
|
Automatically calculates X, SD, CV% and saves date
|
Language
|
English and Spanish. Other language on request
|
Power supply
|
110/220V±10% , 60/50Hz
|
Package Size
|
660*480*640mm
|
GrossWeight
|
38kg
|